Why the study?
Does the Rotem Platelet system provide comparable platelet reactivity measurements to the Multiplate system in cardiac surgery patients?
Population
50 cardiac surgery patients
Comparison
Preoperative platelet function testing with… vs Preoperative platelet function testing with…
Design
Cross-sectional
Authors
Loading...
Rotem Platelet and Multiplate ADPtest results lack agreement; leaves open validation of either for post-P2Y12 assessment in cardiac surgery.
Does the Rotem Platelet system provide comparable platelet reactivity measurements to the Multiplate system in cardiac surgery patients?
Platelet function assessment with Rotem Platelet greatly differs from Multiplate, with low precision preventing the use of a correction value, and Multiplate being more reliable for assessing reactivity after P2Y12 inhibitor discontinuation.
Ranucci et al. (2015) studied this question.