An amplifier that is suitable for use in a variety of nuclear-particle counting experiments is described. The high and low frequency response of the amplifier can be varied by means of a three-position switch. In the wide-band position the transient response to a step function is a voltage pulse with a rise time of 0.15 μsec. and a total duration of less than 0.5 μsec. In the narrow-band position a signal of 3 microvolts produces a voltage pulse at the output equal to r.m.s. noise. A pulse-height selector is included for measuring pulses up to 100-volts amplitude.