A spectrofluorimetric method for the determination of ultratrace amounts of lead and cadmium is described based on the sequential extraction of the ternary ion-association complexes formed between the cation, a cryptand as the ligand and eosin as a counter ion. A linear working range from the detection limit (0.5 ng ml(-1) to 250 ng ml(-1) of lead and to 1 50 ng ml(-1) of cadmium was obtained. The relative standard deviation was 2-4%. The proposed method has been applied successfully to the determination of lead and cadmium in zinc metals and soft drinks.
No takes yet. Share an insight, caveat, or question.
Gomis et al. (1990) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: