Answers are grounded in this paper and cite their sources.
. However, to be used effectively for magnetoencephalography (MEG), dense arrays of these sensors are required to operate as an integrated turn-key system. In this study, we present the HEDscan, a 128-sensor OPM MEG system by FieldLine Medical, and evaluate its sensor performance with regard to bandwidth, linearity, and crosstalk. We report results from cross-validation studies with conventional cryogenic MEG, the Magnes 3,600 WH Biomagnetometer by 4-D Neuroimaging. Our results show high signal amplitudes captured by the OPM-MEG system during a standard auditory paradigm, where short tones at 1000 Hz were presented to the left ear of six healthy adult volunteers. We validate these findings through an event-related beamformer analysis, which is in line with existing literature results.
No takes yet. Share an insight, caveat, or question.
Alem et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: