Answers are grounded in this paper and cite their sources.
We have been developing a superconducting time-of-flight mass spectrometry (TOF-MS) system that consists of a superconducting strip ion detector (SSID) and a single-flux-quantum (SFQ) time-to-digital converter. In this study, we implement a prototype TOF-MS system using an SSID and an SFQ readout circuit in which output signals from the SSID are read out by the SFQ readout circuit and output to room-temperature electronics. The SFQ readout circuit, which consists of a current discriminator, a Josephson transmission line and an SFQ/dc converter, was fabricated using the AIST Nb standard process (STP2), and installed in a 4.2 K cryostat with a meander-shaped NbN SSID measuring 200 μ m square. The dark count rate for the SSID was measured as increasing exponentially with the increase in the bias current of the SSID by using the SFQ readout circuit. Mass spectrum measurements of biomolecules, Angiotensin I, which has a molecular weight of 1296 Da, were demonstrated by using the matrix-assisted laser desorption/ionization method, and a clear corresponding peak was observed in the mass spectrum.
No takes yet. Share an insight, caveat, or question.
Sano et al. (2015) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: