Key result
PEGME homogeneous HDL-C assay outperforms PPD assay for routine testing with ~0.99 correlation to reference.
Answers are grounded in this paper and cite their sources.
We evaluated the performance of two homogeneous assays for quantifying HDL cholesterol (HDL-C) and compared them with the phosphotungstic acid (PTA)/ MgCl2 assay. Both homogeneous HDL-C assays were precise, having a within-run CV of < 1.20% and a between-run CV of < 4.07%. The HDL-C values (y) measured by the two homogeneous methods correlated well with those by the PTA/MgCl2 method (x): y = 1.00x + 64.98 mg/L, r = 0.987, Sy/x = 27.99 mg/L (n = 152) for the polyethylene glycol-modified enzymes/alpha-cyclodextrin sulfate (PEGME) assay (Kyowa), and y = 0.84x + 106.51 mg/L, r = 0.984, Sy/x = 26.10 mg/L (n = 152) for the polyanion-polymer/detergent (PPD) assay (Daiichi). The specificity of the PEGME method seemed better than that of the PPD method, as the PPD method was markedly interfered with by supplemental LDL-C. Addition of 20 g/L triglycerides produced a negative error of approximately 18% in both homogeneous assays. Bilirubin and hemoglobin had little influence on the PEGME method; hemoglobin had little effect on the PPD method. Bilirubin, however, markedly decreased the readings by the PPD method. We found the PEGME assay superior to the PPD assay for routine HDL-C testing, because the PPD assay is relatively inaccurate and not specific.
No takes yet. Share an insight, caveat, or question.
Huang et al. (1997) studied this question. Homogeneous HDL-C assays (PEGME and PPD methods) vs. Phosphotungstic acid (PTA)/MgCl2 assay was evaluated on Correlation and analytical performance of HDL-C quantification compared to PTA/MgCl2 method (r = 0.987 (PEGME), r = 0.984 (PPD)). The PEGME homogeneous HDL-C assay was superior to the PPD assay for routine testing, with both correlating well with the PTA/MgCl2 method.
Synapse has enriched 10 related papers. Consider them for comparative context: