An electron absorption spectrometer with an improved velocity analyzer using the principle of individually designed stages for deceleration, dispersion, and reacceleration is described. The instrument is usable over an energy range of 1 to 50 kev and can obtain angular distribution data over a range of ±0.1 radian. The angular resolution is better than 10−3 radian. The demonstrated energy resolution at 20 kev is 0.6 ev. This resolution appears to be limited by the primary beam energy spread which is characteristic of a beam produced by a telefocus gun with a tungsten cathode.