Does the self-expandable ACURATE TA valve system improve clinical outcomes compared to the balloon-expandable SAPIEN XT valve system in patients undergoing transapical TAVI?
Does the self-expandable ACURATE TA valve system improve clinical outcomes compared to the balloon-expandable SAPIEN XT valve system in patients undergoing transapical TAVI?
The ACURATE TA and SAPIEN XT transapical TAVI devices demonstrated comparable clinical and hemodynamic outcomes, though the ACURATE valve required more frequent post-ballooning.
Comparable outcomes in matched transapical TAVI; hypothesis-generating for post-ballooning differences and requires randomized confirmation.
OBJECTIVES: Several alternative transapical (TA) aortic prostheses for aortic valve implantation (AVI) have recently become available. Data directly comparing the performance of these different TA-AVI valves, however, are sparse. Therefore, we sought to analyse the performance of the better-established balloon-expandable SAPIEN XT™ valve system, and compare it with the recently approved self-expandable ACURATE TA™ system by means of propensity score (PS) analysis. METHODS: Samples from a total of 310 consecutive patients treated with transapical transcatheter aortic valve implantation (TAVI) were included (2010-2014). The ACURATE valve was implanted in 40% (n = 131), SAPIEN in 60% (n = 179). To adjust for baseline differences, 1:1 PS matching was used, and resulted in 103 matched pairs. In addition to demographical and clinical variables, potential anatomical confounders (calcification grade: computed tomography (CT) Agatston score; annulus diameter: CT-effective diameter) were included within the PS estimation. RESULTS: For ACURATE- versus SAPIEN-treated patients, the mean age was 83.1 ± 5.4 vs 79.8 ± 9.5 (P < 0.001). Mean Society of Thoracic Surgeon scores were 7.7 ± 4.6 vs 8.1 ± 6.0 (P = 0.56). PS matching resulted in bias reduction <0.2, indicating excellent balance and allowing for valid outcome comparison. Device success, 30-day mortality rate and 1-year survival were comparable. Pacemaker and 30-day neurological event rates were not significantly different. The need for post-ballooning was significantly higher in the self-expandable group (40% vs 9%; P < 0.001). Echocardiography upon discharge demonstrated excellent functional outcomes regarding residual paravalvular leaks (ACURATE: 3% vs SAPIEN: 4%, P = 1.0) with slightly higher mean gradients in the ACURATE group (12 ± 5 mmHg vs 10 ± 5 mmHg, P = 0.003). CONCLUSIONS: The two transapical TAVI devices demonstrated comparable haemodynamic performance and clinical outcome. The self-expandable valve required more frequent post-ballooning without affecting the safety profile.
No takes yet. Share an insight, caveat, or question.
Kempfert et al. (2015) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: