Why the study?
Automated ECG interpretations may be erroneous and lead to erroneous physician overreads, including for AF.
Does the Cardiologs deep neural network algorithm improve the accuracy of atrial fibrillation diagnosis on 12-lead ECGs compared to the conventional Veritas algorithm and physician overread?
Does the Cardiologs deep neural network algorithm improve the accuracy of atrial fibrillation diagnosis on 12-lead ECGs compared to the conventional Veritas algorithm and physician overread?
A deep neural network algorithm (Cardiologs) outperformed a conventional algorithm and matched physician overread accuracy for diagnosing atrial fibrillation on 12-lead ECGs.
May support Cardiologs for AF detection on ECGs without overread; leaves open prospective validation before clinical adoption.
Automated electrocardiogram (ECG) interpretations may be erroneous, and lead to erroneous overreads, including for atrial fibrillation (AF). We compared the accuracy of the first version of a new deep neural network 12-Lead ECG algorithm (Cardiologs®) to the conventional Veritas algorithm in interpretation of AF. 24,123 consecutive 12-lead ECGs recorded over 6 months were interpreted by 1) the Veritas® algorithm, 2) physicians who overread Veritas® (Veritas® + physician), and 3) Cardiologs® algorithm. We randomly selected 500 out of 858 ECGs with a diagnosis of AF according to either algorithm, then compared the algorithms' interpretations, and Veritas® + physician, with expert interpretation. To assess sensitivity for AF, we analyzed a separate database of 1473 randomly selected ECGs interpreted by both algorithms and by blinded experts. Among the 500 ECGs selected, 399 had a final classification of AF; 101 (20.2%) had ≥1 false positive automated interpretation. Accuracy of Cardiologs® (91.2%; CI: 82.4–94.4) was higher than Veritas® (80.2%; CI: 76.5–83.5) (p < 0.0001), and equal to Veritas® + physician (90.0%, CI:87.1–92.3) (p = 0.12). When Veritas® was incorrect, accuracy of Veritas® + physician was only 62% (CI 52–71); among those ECGs, Cardiologs® accuracy was 90% (CI: 82–94; p < 0.0001). The second database had 39 AF cases; sensitivity was 92% vs. 87% (p = 0.46) and specificity was 99.5% vs. 98.7% (p = 0.03) for Cardiologs® and Veritas® respectively. Cardiologs® 12-lead ECG algorithm improves the interpretation of atrial fibrillation.
No takes yet. Share an insight, caveat, or question.
Smith et al. (2019) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: