Answers are grounded in this paper and cite their sources.
We report the use of a programmable array microscope (PAM) for the acquisition of spectrally resolved and high-throughput optical sections. The microscope is based on the use of a spatial light modulator for defining patterns of excitation and/or detection of fluorescence. For obtaining optically sectioned spectral images, the entrance slit of an imaging spectrograph and a line illumination pattern defined with a spatial light modulator are placed in conjugate optical positions. Compared to wide-field illumination, optical sectioning led to greater than 3× improvement in the rejection of out-of-focus fluorescence emission and nearly 6× greater peak-to-background ratios in biological specimens, yielding better contrast and spectral characterization. These effects resulted from a reduction in the artifacts arising from spectral contributions of structures outside the region of interest. We used the programmable illumination capability of the spectroscopic system to explore a variety of excitation/detection patterns for increasing the throughput of optical sectioning microscopes. A Sylvester-type Hadamard construction was particularly efficient, performing optical sectioning while maintaining a 50% optical throughput. These results demonstrate the feasibility of full-field highly multiplexed confocal spectral imaging.
No takes yet. Share an insight, caveat, or question.
Hanley et al. (1998) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: