Key result
Motion-corrected spatiotemporal clutter filtering improved the contrast-to-noise ratio of thyroid small vessel blood flow compared to standard SVD (8.17 and 8.32 dB vs 0.87 and 4.33 dB).
Why the study?
Non-invasive assessment of thyroid small vessel blood flow without contrast agents is challenging due to physiological motion from the adjacent pulsating carotid artery.
Does motion correction prior to SVD-based clutter filtering improve visualization of small vessel blood flow in the thyroid compared to standard SVD?
Does motion correction prior to SVD-based clutter filtering improve visualization of small vessel blood flow in the thyroid compared to standard SVD?
Motion correction prior to SVD-based clutter filtering improves the visualization of small vessel blood flow in the thyroid without contrast agents.
May facilitate non-contrast thyroid microvascular imaging; leaves open diagnostic utility pending clinical validation.
Reliable assessment of small vessel blood flow in the thyroid, without using any contrast agents, can be challenging because of increased physiological motion resulting from its proximity to the pulsating carotid artery. In this study, we hypothesized that correction of tissue motion prior to singular value decomposition (SVD)-based clutter filtering can improve the coherency of the tissue components and, thus, may allow better clutter suppression and visualization of small vessels in the thyroid. We corroborated this hypothesis by conducting phantom and in vivo studies using a clinical ultrasound scanner implemented with compounded plane wave imaging. The phantom studies were conducted using a homogeneous tissue-mimicking phantom to study the impact of motion on the covariance of the spatiotemporal Doppler data, in the absence of blood activity. The non-invasive in vivo study was conducted on a 74-y-old woman with a thyroid nodule suspicious of malignancy. A rigid body-based motion correction was performed using tissue displacements obtained from 2-D normalized cross-correlation-based speckle tracking. Subsequently, the power Doppler images were computed using SVD-based spatiotemporal clutter filtering. The results from the phantom study revealed that motion can considerably reduce the covariance of the spatiotemporal data and, thus, increase the rank of the tissue components. When the phantom was subjected to a total translation displacement of 6 pixels over the entire ensemble, in each direction (axial and lateral), the covariance dropped by more than 25%. The results obtained from the non-invasive in vivo study indicated that visualization of small vessel blood flow improved with motion correction of the power Doppler ensemble. The contrast-to-noise ratio of the blood signal in motion-corrected power Doppler images was considerably higher (8.17 and 8.32 dB), compared with that obtained using the standard SVD approach at an optimal threshold (0.87 and 4.33 dB) and a lower singular value threshold (1.92 and 3.05 dB). Further, the covariance of the in vivo thyroid spatiotemporal data increased by approximately 10% with motion correction. These preliminary results indicate that motion correction can be used to improve the visualization of small vessel blood flow in the thyroid, without using any contrast agents. The results of this feasibility study were encouraging, and warrant further development and more in vivo validation in moving tissues and organs.
No takes yet. Share an insight, caveat, or question.
Nayak et al. (2019) studied Thyroid nodule suspicious of malignancy (n=1). Motion-corrected spatiotemporal clutter filtering vs. Standard SVD approach was evaluated on Contrast-to-noise ratio of the blood signal. Motion-corrected spatiotemporal clutter filtering improved the contrast-to-noise ratio of thyroid small vessel blood flow compared to standard SVD (8.17 and 8.32 dB vs 0.87 and 4.33 dB).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: