This is the authors' abstract. We don't add key points for this paper.
The use of a sputter-cleaned copper/silver/gold (Cu/Ag/Au) alloy for calibrating the energy scale and instrument resolution (as full width at half-maximum or FWHM) of x-ray photoelectron spectrometers (XPS) was evaluated. The three component alloy material is a convenient alternative to the common practice of using three separate sputter-cleaned reference materials (i.e., Cu, Ag, and Au foils) for the calibration of XPS. The Cu 3p3/2, Cu 2p3/2, Cu L3M4,5M4,5, Ag 3d5/2, Ag M4NN, and Au 4f7/2 peaks have been recommended in the literature for calibrating XPS spectrometers. Results indicate that the Cu/Ag/Au alloy yields XPS binding energies for these peaks within reasonable experimental error (i.e., ± 0.1–0.2 eV or less) of the values obtained for pure Cu, Ag, and Au materials. In addition, the FWHM values obtained for these peaks with the Cu/Ag/Au alloy were essentially equivalent (within 0–0.03 eV) with the values obtained for pure Cu, Ag, and Au foils.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Brian R. Strohmeier (1994) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: