Key result
Novel triple-row rotator cuff repairs resulted in significantly higher ultimate failure loads (465 N) compared to single-row (414 N) and double-row (394.5 N) repairs in a porcine model.
Why the study?
Does a novel triple-row rotator cuff repair improve biomechanical properties compared to single- or double-row repairs in a porcine model?
Does a novel triple-row rotator cuff repair improve biomechanical properties compared to single- or double-row repairs in a porcine model?
Absolute Event Rate: 465% vs 414%
p-value: p=0.0084
In a porcine model, a novel triple-row rotator cuff repair demonstrated superior biomechanical properties, including smaller gap formation and higher ultimate failure load, compared to single- and double-row repairs.
Does not support clinical adoption; hypothesis-generating for triple-row techniques in human rotator cuff repair.
Comparison of a novel triple-row (NTR) rotator cuff repair (RCR) was made to single row (SR) and double row (DR) repairs as measured by gap formation, Ultimate Failure Load (UFL), and mode of failure. Porcine humeri were divided into three groups of forty and the infraspinatus tendons were completely released and then repaired using generic (PEEK) anchors. Biomechanical testing of RCRs consisted of 3500 cycles of a load varied from 10 to 180 Newtons (N). Gap size at all cycle counts was significantly larger for SR and DR than NTR RCRs. 84% of NTR repairs completed the 3500 cycle test versus 66% of DR and 61% of SR repairs. The tendon pulled from the sutures 60%, 49% and 33% of the time for SR, DR and NTR RCRs respectively. UFLs were 464 N for the NTR repairs, 394 N for DR and 414 N for SR repairs. In summary, NTR repairs resulted in smaller gap formation, higher UFLs, and fewer instances of tendons tearing at the sutures than SR or DR repairs. Results of NTR repairs demonstrate that the tendon footprint is securely restored with small gap formation for the range of forces that a shoulder will be subjected to in normal use.
No takes yet. Share an insight, caveat, or question.
Blair A Rhode (2015) studied Rotator cuff tear (porcine model) (n=120). Novel triple-row (NTR) rotator cuff repair vs. Single-row (SR) and double-row (DR) rotator cuff repairs was evaluated on Ultimate Failure Load (UFL) (p=0.0084). Novel triple-row rotator cuff repairs resulted in significantly higher ultimate failure loads (465 N) compared to single-row (414 N) and double-row (394.5 N) repairs in a porcine model.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: