This is the authors' abstract. We don't add key points for this paper.
Flow injection (FI) has been applied to the determination of CrIII in water and food samples. The method is based on measurement of the light emitted from the CrIII-catalysed oxidation of luminol by H2O2. The apparatus consists of an FI system with a flow cell suitable for chemiluminescence detection. The flow cell, situated near the photodetector, is a coiled tube made from transparent polytetrafluoroethylene. The typical signal is a narrow peak, the height of which is proportional to the light emitted and, therefore, to the concentration of CrIII. The detection limit is 0.01 ppb and the linear range extends up to 6 ppb. The concentrations of the reagents, the pH and the flow rates were optimized. Interferences by several metal ions were examined, and the system was shown to possess a high selectivity. The method was successfully applied to the determination of CrIII in water and food samples.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Escobar et al. (1993) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: