A high-resolution Thomson scattering diagnostic is in operation on the Alcator C-Mod tokamak, measuring radial profiles of electron temperature and density at the plasma edge. Photons are scattered from a Nd–yttrium–aluminum–garnet laser beam pulsed at 30 Hz (1.3 J, 8 ns pulse), and are measured by a filter polychromator with four spectral channels. The polychromator measures Te in the range of 15–800 eV and ne of 0.3–3×1020 m−3. Twenty scattering volumes are located about the last closed flux surface, spaced for a nominal resolution of 1.3 mm in midplane radial coordinates. High resolution is essential for measuring edge Te and ne profiles on C-Mod, since these quantities exhibit gradient scale lengths as small as 2 mm in H mode. The steep profiles at the H mode edge are fit to a parameterized pedestal function for ease of analysis. Measured profiles are compared with edge profiles from electron cyclotron emission and visible continuum diagnostics.