Why the study?
The authors sought to investigate the defibrillation safety margin and the efficacy of terminating induced ventricular fibrillation at S-ICD implantation using energies lower than the recommended 65 J.
Does low-energy shock testing (20-50 J) effectively terminate induced ventricular fibrillation in S-ICD recipients?
Comparison
Induced VF termination attempts using lower energies (20 to 50 J) vs recommended 65 J
Authors
Loading...
Low-energy VF termination rates support cautious S-ICD energy optimization; leaves open need for randomized safety-margin trials.
Does low-energy shock testing (20-50 J) effectively terminate induced ventricular fibrillation in S-ICD recipients?
Low-energy shocks (≤40 J) successfully terminate induced VF in 84% of S-ICD recipients, suggesting that a 60 J output device may provide an adequate safety margin while enabling device downsizing.
Biffi et al. (2019) studied this question.