Why the study?
Does routine DFT testing at the time of ICD implant improve 5-year survival in patients receiving ICDs?
Does routine DFT testing at the time of ICD implant improve 5-year survival in patients receiving ICDs?
Routine DFT testing during ICD implantation provides a marginal and non-significant 5-year survival benefit compared to no testing in a decision analysis model.
Answers are grounded in this paper and cite their sources.
Supports deferring routine DFT testing at ICD implantation; leaves open the need for randomized confirmation.
OBJECTIVE: To assess the impact of defibrillation threshold (DFT) testing of implanted cardioverter-defibrillators (ICDs) on survival. BACKGROUND: DFT testing is generally performed during implantation of ICDs to assess sensing and termination of ventricular fibrillation. It is common clinical practice to defibrillate ventricular fibrillation twice at an output at least 10 J below the maximum output of the device, providing a 10 J safety margin. However, there are few data regarding impact of DFT testing on outcomes. METHODS: Decision analysis and Monte Carlo simulation were used to assess expected outcomes of DFT testing. Survival of a hypothetical cohort of patients was assessed according to two strategies-routine DFT testing at time of ICD implant versus no DFT testing. Assumptions in the model were varied over a range of reasonable values to assess outcomes under a variety of scenarios. RESULTS: Five-year survival with DFT and no-DFT strategies were similar at 59.72% and 59.36%, respectively. The results were not sensitive to changing risk estimates for arrhythmia incidence and safety margin. Results of the Monte Carlo simulation were qualitatively similar to the base case scenario and consistent with a small and nonsignificant survival advantage with routine DFT testing. CONCLUSIONS: The impact of DFT testing on 5-year survival in ICD patients, if it exists, is small. Survival appears higher with DFT testing as long as annual risk of lethal arrhythmia or the risk of a narrow safety margin is at least 5%, although the incremental benefit is marginal and 95% confidence intervals cross zero. A prospective randomized study of DFT testing in modern devices is warranted.
No takes yet. Share an insight, caveat, or question.
Gula et al. (2008) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: