Subcutaneous fat beneath the lead and generator can challenge defibrillation threshold testing, raising concerns about extravascular ICD performance in patients with obesity.
Is extravascular implantable cardioverter-defibrillator (EV-ICD) implantation feasible and effective in achieving successful defibrillation threshold testing in patients with class III obesity (BMI > 45 kg/m²)?
Is extravascular implantable cardioverter-defibrillator (EV-ICD) implantation feasible and effective in achieving successful defibrillation threshold testing in patients with class III obesity (BMI > 45 kg/m²)?
Extravascular ICD implantation demonstrated favorable defibrillation threshold testing (30 J) and low high-voltage impedance in two patients with class III obesity (BMI > 45 kg/m²), suggesting it is a promising alternative to subcutaneous ICDs in this population.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for EV-ICD in class III obesity; larger prospective studies needed before practice change.
One of the challenges with the subcutaneous implantable cardioverter-defibrillator (S-ICD) is the amount of subcutaneous fat beneath the lead and generator during implantation, which can affect defibrillation threshold (DFT) testing. The extravascular ICD (EV-ICD), created to address many of the limitations of the S-ICD, carried a similar concern given its external placement. Importantly, in the EV-ICD Pivotal trial there was no statistical difference in implant DFT success by body mass index (BMI), height, or weight.