Key result
S-ICD cuts lead-related complications and invasive interventions vs transvenous ICD.
Why the study?
The S-ICD was developed to overcome TV-ICD lead complications and systemic infections, prompting this prespecified secondary analysis to evaluate all complications in the PRAETORIAN trial.
Does the subcutaneous ICD reduce device-related complications and need for invasive interventions compared to transvenous ICD in patients with an indication for ICD therapy?
RCT (n=849)
Randomized
Yes
Does the subcutaneous ICD reduce device-related complications and need for invasive interventions compared to transvenous ICD in patients with an indication for ICD therapy?
S-ICD therapy is associated with fewer lead-related complications, systemic infections, and complications requiring invasive interventions compared to TV-ICD therapy over a median follow-up of 49 months.
Answers are grounded in this paper and cite their sources.
S-ICD reduces invasive interventions and lead complications versus TV-ICD; confirms safety advantage for device selection in RCT.
BACKGROUND: The subcutaneous implantable cardioverter-defibrillator (S-ICD) is developed to overcome lead-related complications and systemic infections, inherent to transvenous ICD (TV-ICD) therapy. The PRAETORIAN trial demonstrated that the S-ICD is non-inferior to the TV-ICD with regard to the combined primary endpoint of inappropriate shocks and complications. This prespecified secondary analysis evaluates all complications in the PRAETORIAN trial. METHODS AND RESULTS: The PRAETORIAN trial is an international, multicentre, randomized trial in which 849 patients with an indication for ICD therapy were randomized to receive an S- ICD (N = 426) or TV-ICD (N = 423) and followed for a median of 49 months. Endpoints were device-related complications, lead-related complications, systemic infections, and the need for invasive interventions. Thirty-six device-related complications occurred in 31 patients in the S-ICD group of which bleedings were the most frequent. In the TV-ICD group, 49 complications occurred in 44 patients of which lead dysfunction was most frequent (HR: 0.69; P = 0.11). In both groups, half of all complications were within 30 days after implantation. Lead-related complications and systemic infections occurred significantly less in the S-ICD group compared with the TV-ICD group (P < 0.001, P = 0.03, respectively). Significantly more complications required invasive interventions in the TV-ICD group compared with the S-ICD group (8.3% vs. 4.3%, HR: 0.59; P = 0.047). CONCLUSION: This secondary analysis shows that lead-related complications and systemic infections are more prevalent in the TV-ICD group compared with the S-ICD group. In addition, complications in the TV-ICD group were more severe as they required significantly more invasive interventions. This data contributes to shared decision-making in clinical practice.
No takes yet. Share an insight, caveat, or question.
Knops et al. (2022) conducted an RCT in Indication for ICD therapy (n=849). Subcutaneous implantable cardioverter-defibrillator (S-ICD) vs. Transvenous implantable cardioverter-defibrillator (TV-ICD) was evaluated on Device-related complications (HR 0.69, p=0.11). Subcutaneous ICD reduced lead-related complications, systemic infections, and complications requiring invasive interventions vs transvenous ICD.
Synapse has enriched 10 related papers. Consider them for comparative context: