A way of using gravity flow to induce a linear convection within a microfluidic system is presented. It is shown and mathematically supported that tilting a 1 cm long covered microchannel is enough to generate flow rates up to 1000 nL.min(-1), which represents a linear velocity of 2.4 mm.s(-1). This paper also presents a method to monitor the microfluidic events occurring in a covered microchannel when a difference of pressure is applied to force a solution to flow in said covered microchannel, thanks to electrodes inserted in the microfluidic device. Gravity-induced flow monitored electrochemically is applied to the performance of a parallel-microchannel enzyme-linked immunosorbent assay (ELISA) of the thyroid-stimulating hormone (TSH) with electrochemical detection. A simple method for generating and monitoring fluid flows is described, which can, for instance, be used for controlling parallel assays in microsystems.
No takes yet. Share an insight, caveat, or question.
Morier et al. (2004) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: