Theoretical calculations of the polychromatic modulation transfer function (MTF) and wave-front aberration were performed with physiological eye models. These eye models have an amount of spherical aberration that is representative of a normal population of pseudophakic eyes implanted with two different types of intraocular lens (IOL) made from high-refractive-index silicone. These theoretical calculations were compared with the measured contrast sensitivity function (CSF) under mesopic lighting conditions and with wave-front aberration (obtained with a Hartmann-Shack wave-front sensor) collected from 37 patients bilaterally implanted with the same types of lens. The relationships between the ocular wave-front aberration and the MTF predicted by the eye models and the CSF and the ocular wave-front aberration measured in eyes implanted with IOLs were investigated. The predicted improvements in MTF and wave-front aberration correlated well with the improvements measured in practice. Physiological eye models are therefore useful tools for IOL design.