A fully automated on-line reversed phase liquid chromatography-gas chromatography system is described. The system uses a prototype of the automated through oven transfer adsorption desorption interface. The system is demonstrated by presenting a new rapid method for the determination of pesticide residue in olive oil, which is injected directly with no sample pretreatment step other than filtration. Methanol:water is used as the eluent in the LC preseparation step, while the LC fraction containing the pesticide is automatically transferred to the gas chromatograph. Detection limits of pesticides varied from 0.18 to 0.44 mg/L when a flame ionization detector was used. As an example, relative standard deviation and linear calibration are presented for terbutryne.