This is the authors' abstract. We don't add key points for this paper.
Simultaneous measurement of temperature and pressure with a single fiber Bragg grating (FBG) based on a broadened reflection spectrum is proposed and experimentally demonstrated. A novel double-hole structure of a cantilever beam is designed, and a FBG is affixed on the nonuniform strain area of the cantilever beam. The Bragg reflection bandwidth is sensitive to the spatially gradient strain but is free from the spatially uniform temperature. The wavelength peak shift and the bandwidth broadening of the FBG with a change of temperature and pressure allow for simultaneous discrimination between the temperature and the pressure effects. Standard deviation errors of 1.4 degrees C and 1.8 kPa were obtained with temperature and pressure ranges of 20 degrees C-100 degrees C and 0-80 kPa, respectively. This novel and low-cost sensor approach has considerable potential applications for temperature-insensitive strain measurement.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Guo et al. (2006) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: