PURPOSE: To demonstrate in-vivo imaging with a low-cost, low-field MRI scanner built and operated in Africa, and to show how systematic hardware and software improvements can mitigate the main operational limitations encountered in low-resource environments. METHODS: A 46 mT Halbach scanner located at the Mbarara University of Science and Technology (Uganda) was upgraded through a complete reorganization of grounding and shielding, installation of new control electronics and open-source user-interface software. Noise performance was quantified using a standardized protocol and in-vivo brain images were acquired with three-dimensional RARE sequences. Distortion correction was implemented using cloud-based reconstructions incorporating magnetic field maps. RESULTS: The revamped system reached noise levels routinely below three times the thermal limit and demonstrated stable operation over multi-day measurements. Three-dimensional T1- and T2-weighted brain images were successfully acquired and distortion-corrected with remote GPU-based reconstructions and near real-time visualization through the user interface. CONCLUSIONS: The results show that low-cost MRI systems can achieve clinically relevant image quality when electromagnetic noise and power-grid instabilities are properly addressed. This work highlights the feasibility of sustainable MRI development in low-resource settings and identifies stable power delivery and local capacity building as the key next steps toward clinical translation.