Answers are grounded in this paper and cite their sources.
The determination of intra-abdominal pressures (IAP) may be useful in many clinical situations. The authors recently demonstrated in the canine model a close correlation between actual IAP and the bladder pressure measurements obtained from a transurethral catheter. The purpose of this study was to clinically validate this technique. The authors studied 16 patients, and compared IAP in three positions (supine, with compressions, and semi-erect) utilizing both direct intraperitoneal pressure monitoring and the pressure obtained with a transurethral bladder catheter. Their results demonstrated a linear relationship between the two methods described, with a mean r value of 0.95 in the supine and semi-erect positions, and 0.99 with abdominal compressions (P less than 0.0001). The authors conclude that transurethral monitoring of bladder pressure offers a safe, simple, and highly accurate method for evaluating IAP at the bedside. Studies evaluating the indication for its use in the operating room and intensive care settings appear warranted.
No takes yet. Share an insight, caveat, or question.
Iberti et al. (1989) studied this question.