A method for obtaining histograms of the frequency of occurrence of optical densities in mammograms has been implemented. Averaged optical density distributions were produced for approximately 50 screening mammograms at each of two centres using different film-screen and processing chemistry combinations. Additional measurements were made on films of two commercially available anthropomorphic breast phantoms. The results show that screening mammograms, on average, exhibit a strongly peaked optical density distribution which corresponds closely to the target optical density measured using the current recommended test protocol for automatic exposure control devices.
No takes yet. Share an insight, caveat, or question.
Kotre et al. (1994) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: