In 2012, we presented the Hyperion-IIDpreclinical PET insert which uses Philips Digital Photon Counting's digital SiPMs and is designed to be operated in a 3-T MRI. In this work we use the same platform equipped with scintillators having dimensions closer to a clinical application. This allows an investigation of the time of flight (ToF) performance of the platform and its behavior during simultaneous MR operation. We employ LYSO crystal arrays of 4×4 ×10 mm3coupled to 4 ×4 PDPC DPC 3200-22 sensors (DPC) resulting in a one-to-one coupling of crystals to read-out channels. Six sensor stacks are mounted onto a singles processing unit in a 2 ×3 arrangement. Two modules are mounted horizontally facing each other on a gantry with a crystal-to-crystal spacing of 217.6 mm (gantry position). A second arrangement places the modules at the maximum distance of approximately 410 mm inside the MR bore (maximum distance position) which brings each module close to the gradient system. The DPCs are cooled down to approximately 5-10° C under operation. We disable 20% of the worst cells and use an overvoltage of Vov= 2.0 V and 2.5 V. To obtain the best time stamps, we use the trigger scheme 1 (first photon trigger), a narrow energy window of 511 ±50 keV and a minimum required light fraction of the main pixel of more than 65% to reject intercrystal scatter. By using a22Na point source in the isocenter of the modules, the coincidence resolution time (CRT) of the two modules is evaluated inside the MRI system without MR activity and while using highly demanding gradient sequences. Inside the B0field without any MR activity at an overvoltage of Vov= 2.0 V, the energy resolution is 11.45% (FWHM) and the CRT is 250 ps (FWHM). At an overvoltage of Vov= 2.5 V, the energy resolution is 11.15% (FWHM) and the CRT is 240 ps (FWHM). During a heavy z-gradient sequence (EPI factor: 49, gradient strength: 30 mT/m, slew rate: 192.3 mT/m/ms, TE/TR: 12/25 ms and switching duty cycle: 67%) at the gantry position and an overvoltage of Vov= 2.0 V, the energy resolution is degraded relatively by 4.1% and the CRT by 25%. Using the same sequence but at the maximum distance position and an overvoltage of Vov= 2.5 V, we measure a degradation of the energy resolution of 9.2% and a 52% degradation of the CRT. The Hyperion-IIDplatform proofs to deliver good timing performance and energy resolution inside the MRI system even under highly demanding gradient sequences.
No takes yet. Share an insight, caveat, or question.
Schug et al. (2015) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: