Systematic voltammetric and amperometric studies have been undertaken to examine the electrochemical behaviour of uric acid at bare and surface-modified screen-printed electrodes. The precision of the electrode manufacture was determined by cyclic voltammetry with a 1.0 × 10–4 mol dm–3 uric acid solution and was calculated to be 6.0%(n= 5). Several strategies were investigated in an attempt to eliminate interference from ascorbic acid. These involved coating the electrode surface with Nafion, or the enzyme L-ascorbic acid oxidase. The latter was immobilized using one of two methods: either by a simple adsorption process, or by cross-linking with bovine serum albumin and glutaraldehyde. The amperometric response at the surface-adsorbed enzyme electrode for uric acid was linear over the concentration range from 5.08 × 10–6 to 1.51 × 10–4 mol dm–3; the limit of detection was 2.54 × 10–6 mol dm–3 using a full-scale deflection of 0.5 µA. This modified electrode exhibited no response to ascorbic acid at levels up to 0.53 mmol dm–3. The electrode modified by cross-linking the enzyme to the surface showed no response to ascorbic acid concentrations of up to 0.093 mmol dm–3.
No takes yet. Share an insight, caveat, or question.
Gilmartin et al. (1992) studied this question.
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: