Key result
Novel lithium dilution method accurately measures cardiac output and shunt flow in venovenous ECMO.
Why the study?
A method to measure cardiac output and shunt fraction during venovenous ECMO was needed to improve patient management and cannula design.
Does a lithium dilution method accurately measure cardiac output and shunt fraction in an in vitro venovenous ECMO model?
Does a lithium dilution method accurately measure cardiac output and shunt fraction in an in vitro venovenous ECMO model?
The lithium dilution method accurately measures cardiac output and shunt fraction in an in vitro venovenous ECMO model, suggesting potential clinical utility for patient management.
Answers are grounded in this paper and cite their sources.
Supports lithium dilution accuracy in VV ECMO models; leaves open clinical translation pending in vivo validation.
Objective To evaluate, in vitro, a method of measuring cardiac output and shunt fraction during venovenous extracorporeal membrane oxygenation (ECMO). Design Experimental study using an in vitro model. Setting A teaching hospital. Model An ECMO circuit was set up in parallel with a patient circuit consisting of tubing through which saline was circulated from a 50-L reservoir by a pump which was set at 3 L/min to represent cardiac output. A second pump in the ECMO circuit drew saline from the patient circuit and passed it through a membrane oxygenator. The flow from the membrane oxygenator either returned directly to the patient circuit or was diverted, via a third pump, back into the ECMO circuit, thereby producing a shunt. Interventions By adjusting the flow rates of the second (ECMO) and third (shunt) pumps, three shunt fractions of 12%, 25%, and 50% were produced at three different ECMO flow rates. Lithium chloride (0.15 mmol) was injected just downstream of the membrane oxygenator; the lithium ion concentration-time curves were recorded simultaneously in the flow returning to the saline reservoir and in the flow just upstream of the membrane oxygenator using lithium selective electrodes. Measurements and main results Nine pairs of curves were recorded, one pair for each combination of ECMO and shunt flow rates. Analysis of these curves allowed shunt flow and "cardiac output" to be calculated and compared with the flow rates delivered by the pumps. Mean "cardiac output" derived from the lithium dilution curves was 2.98 +/- 0.18 (SD) L/min, compared with a delivered pump flow of 3 L/min. Measured shunt flow = 0.008 + 1.09 x actual shunt flow (R = 0.997). Conclusions This method would allow cardiac output and shunt flow to be measured in patients undergoing venovenous ECMO. It could result in better patient management and improved cannula design.
No takes yet. Share an insight, caveat, or question.
Linton et al. (1998) studied Venovenous extracorporeal membrane oxygenation (ECMO). Lithium dilution method vs. Delivered pump flow rates was evaluated on Cardiac output and shunt fraction (R = 0.997). A new lithium dilution method accurately measured cardiac output (mean 2.98 +/- 0.18 L/min vs actual 3 L/min) and shunt flow (R = 0.997) in an in vitro venovenous ECMO model.