Answers are grounded in this paper and cite their sources.
A flow injection system incorporating a microcolumn of activated alumina was combined with inductively coupled plasma mass spectrometry for on-line trace enrichment and determination of uranium in surface waters and sea water. Deposition of uranium was effected with the use of a basic alumina microcolumn, and injection of nitric acid (250 μL, 2M) served to elute retained species to the ICP. A sample volume of 14 mL resulted in a preconcentration factor of 40, and precision at the 50-ng 1 −1 level was 4.5% (RSD). The procedure was applied to mineral waters, river water, and sea water, and accuracy was assessed through either analysis of certified reference waters or recovery experiments.
No takes yet. Share an insight, caveat, or question.
Dadfarnia et al. (1994) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: