An on-line technique coupling preconcentration via a microprecolumn with microcolumn-liquid chromatography (micro-LC) on a 1-mm inner diameter (i.d.) analytical column was investigated. The system was found to be very simple, fast, and efficient for the trace-level determination of some natural hepatotoxins synthesized by Cyanobacteria, namely, microcystins, in environmental water. Particular attention was paid to band broadening, and a good quality coupling was obtained. A cleanup was introduced in the solid-phase extraction (SPE) sequence to eliminate part of the interferents co- extracted during the sample percolation. The technique was reproducible, with relative standard deviations ranging from 3 to 8%, linear in the range 0.05–50 μg/L. Microcystins could be detected below the 0.04 μg/L level in drinking and surface water using only 5-mL sample volumes, with UV detection. The very small sample volume, resulting from lower detection limits, in absolute amounts, obtained in micro-LC represents a definite advantage over on-line classical chromatography. The suitability of on-line micro-SPE–micro-LC–MS was also evaluated. Application to the determination of microcystins in environmental and drinking water samples are presented.
No takes yet. Share an insight, caveat, or question.
Rivasseau et al. (2000) studied this question.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: