A fully automated anthrax smoke detector (ASD) has been developed and tested. The ASD is intended to serve as a cost effective front-end monitor for anthrax surveillance systems. The principle of operation is based on measuring airborne endospore concentrations, where a sharp concentration increase signals an anthrax attack. The ASD features an air sampler, a thermal lysis unit, a syringe pump, a time-gated spectrometer, and endospore detection chemistry comprised of dipicolinic acid (DPA)-triggered terbium ion (Tb(3+)) luminescence. Anthrax attacks were simulated using aerosolized Bacillus atrophaeus spores in fumed silica, and corresponding Tb-DPA intensities were monitored as a function of time and correlated to the number of airborne endospores collected. A concentration dependence of 10(2)-10(6) spores/mg of fumed silica yielded a dynamic range of 4 orders of magnitude and a limit of detection of 16 spores/L when 250 L of air were sampled. Simulated attacks were detected in less than 15 min.