Key result
Preserved stent circularity after balloon deflation is linked to ~2% lower Melody fracture risk.
Why the study?
Do 3D geometric variations and mechanical stresses during balloon deflation and the cardiac cycle predict stent fracture within 1 year in patients undergoing Melody percutaneous pulmonary valve implantation?
Cohort (n=42)
No
Do 3D geometric variations and mechanical stresses during balloon deflation and the cardiac cycle predict stent fracture within 1 year in patients undergoing Melody percutaneous pulmonary valve implantation?
3D fluoroscopic reconstruction and finite element stress analysis immediately after Melody valve deployment can accurately identify patients at high risk for subsequent stent fracture based on stent circularity loss and compressive recoil stress, potentially enabling tailored follow-up.
Answers are grounded in this paper and cite their sources.
May inform deployment technique to reduce fracture risk; leaves open prospective validation in Melody PPVI.
BACKGROUND: Patients treated with the Melody device (Medtronic) for percutaneous pulmonary valve implantation experience stent fractures in ≈25% of the cases. The aim of this study is to identify the risk factors associated with fracture using 3-dimensional (3D) analyses. METHODS AND RESULTS: In situ 3D shape of the Melody stent was reconstructed from 42 patients using procedural biplane fluoroscopy images, after balloon inflation, at systole and diastole. Four geometric parameters at systole and their variation during balloon deflation and cardiac cycles were measured to describe the 3D strut, cell, section, and stent configuration. Furthermore, patient-specific computer simulations were set up to replicate the history of stent deformations for each patient. Maximum and minimum principal stresses resulting from these analyses were monitored during balloon deflation and cardiac cycle. Univariate logistic regression analyses of 21 geometric parameters and of 4 stress parameters respectively, identified the decreased stent circularity after balloon deflation (odds ratio 0.98; 95% confidence interval, 0.96-0.99; P=0.006) and large compressive stresses during balloon deflation (odds ratio, 0.98; 0.96-0.997; P=0.03), as associated with the risk of fracture. In a multivariable logistic regression model, the 2 covariates identified on univariate analysis (1 geometric and 1 stress) were found to be independently associated with the risk of fracture. The resultant statistical model correctly identified fracture/no fracture in 93% of patients. CONCLUSIONS: Changes in stent section shape after balloon deflation are important variables influencing fracture. This methodology could help design tailored follow-up for patients after percutaneous pulmonary valve implantation.
No takes yet. Share an insight, caveat, or question.
Cosentino et al. (2014) conducted a cohort in Pulmonary artery conduit dysfunction treated with percutaneous pulmonary valve implantation (PPVI) (n=42). 3D stent circularity preservation (change in radial asymmetry during recoil) and compressive stresses during balloon deflation was evaluated on Melody stent fracture within 1 year after percutaneous pulmonary valve implantation (OR 0.98, 95% CI 0.96-0.99, p=0.006). Preserved stent circularity after balloon deflation (OR 0.98, 95% CI 0.96-0.99, P=0.006) and lower compressive stresses (OR 0.98, P=0.03) were associated with reduced Melody stent fracture risk.