The effects on array gain and sidelobe level of a practical digital beamforming (DBF) processor under the wideband conditions typical of ultrasound is discussed. It is concluded that a relatively simple design that replaces each analog delay line with a tapped, digital shift register (DSR) and a digital phase shift operation adjusted for midband will provide the desired performance, provided that the sampling rate of the signal at the input to the DSR is 4 to 10 times the bandwidth. More realistically, when nonidealized passbands are taken into account and the typical condition whereby the transducer frequency is about twice the bandwidth is considered, the rule of thumb for the sampling rate is that it must be 4 to 10 times the transducer frequency.