For the purpose of the quantitative assessment of subtle disease processes in the cardiovascular system an electronically steered sector scanner that combines echographic imaging and Doppler blood velocity measurements has been developed. The integrated operation of a fast Fourier transform (FFT) Doppler signal processor for the simultaneous blood velocity evaluation of 64 individual gates is among the specific design goals. The instrument incorporates an unusually high degree of digital signal processing, which allows for high integration density, easy manufacturing and high reliability in future designs. The complex Doppler spectra are determined for each of the 64 Doppler gates in real time, and the subsequent computation of the first moment provides a reliable estimate of the mean blood flow velocities at the respective locations. The instantaneous velocity profile along the Doppler beam is displayed together with the calculated volume flow rate and a range-selected complete frequency spectrum. Results of both in vitro and in vivo tests indicate that in the future, a higher degree of digital signal processing could be implemented in complex ultrasonic systems.