Using X-ray energies higher than those normally used for macromolecular X-ray crystallography (MX) at synchrotron sources can theoretically increase the achievable signal as a function of dose and reduce the rate of radiation damage [1].In practice, a major stumbling block to the use of higher X-ray energy has been the reduced quantum efficiency of silicon detectors as the X-ray energy increases, but hybrid photon counting CdTe detectors are optimised for higher X-ray energies, and their performance has been steadily improving.The potential advantages of using higher incident beam energy together with a CdTe detector for MX are explored, with a particular focus on the advantages that higher beam energies may have for MX experiments with microbeams or microcrystals.Our Monte Carlo calculations [2] show a greater than a factor of 2 improvement in diffraction efficiency when using microbeams and microcrystals of 5 μm or less.These take into account the escape of photoelectrons from the crystal as well as entry from the surrounding material [3], both of which have now been incorporated into RADDOSE-3D.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Dickerson et al. (2019) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: