Answers are grounded in this paper and cite their sources.
We introduce an apparatus for non-destructive, contact free, reel-to-reel critical current measurement of coated conductor tape. Up to 100 m of tape can be passed in front of a sensor head for the measurement in liquid nitrogen. The tape is transferred and characterized at a speed of 10 mm s−1. After the liquid nitrogen passage it is actively heated up to room temperature to avoid moisture condensation. Two different sensors in the probe head can be optionally activated: a linear Hall probe array in a field coil and an ac-measurement coil picking up third harmonic signals from the film. Once calibrated by transport measurement this arrangement enables characterization of HTS films on magnetic as well as on non-magnetic metal tape substrates. In a 10 mm wide coated conductor, up to 400 A of critical current can be measured.
No takes yet. Share an insight, caveat, or question.
Furtner et al. (2004) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: