Answers are grounded in this paper and cite their sources.
We had successfully operated the 1020-MHz (24.0 T) nuclear magnetic resonance (NMR) spectrometer for 8 months, including the cooling down, shimming, and the 6-month operation for NMR measurements. The NMR superconducting magnet, which had consisted of NbTi and Nb3Sn coils, was upgraded by replacing the innermost Nb3Sn coil with a (Bi,Pb)2Sr2Ca2Cu3O10(Bi-2223) one. The Bi-2223 innermost coil was connected in series with Nb3Sn and NbTi coils. The magnet was cooled with pressurized superfluid helium (He II). The steady-state operating temperature was 1.72 K. The magnet was power-supply driven, i.e., the transport current was always supplied by a power supply. A small normal transition and a power outage occurred during the 8-month operation; however, the magnet quench was avoided.
No takes yet. Share an insight, caveat, or question.
Nishijima et al. (2016) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: