Answers are grounded in this paper and cite their sources.
We demonstrate a potential vision tester (PVT) designed to study and improve the accuracy of visual acuity (VA) measurements in the aging eye. Key features include a high-resolution visual display presented in Maxwellian view, a 3 mm pupil to limit wavefront (WF) aberrations, a Hartmann Shack wavefront sensor to quantify the ocular aberrations, and a deformable mirror to correct optical aberrations. VA was measured using four alternative forced choices for a single black-on-white E stimulus in each trial, with three different levels of correction for the ocular aberrations. For normally sighted subjects, VA was not significantly better when higher-order aberrations beyond second order were corrected.
No takes yet. Share an insight, caveat, or question.
Parimi et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: