A stopped flow temperature jump apparatus is described which permits rapid mixing of reactants and application of a temperature perturbation to the reacting system from 16 msec to several seconds or longer after mixing. The apparatus consists of a flow system with a rectangular observation chamber 1 cm in length. Electrodes are fixed above and below the observation chamber. A stopping syringe activates the time delay, switching, and triggering circuits. The triggering circuit discharges a high voltage capacitor through the solution in the observation chamber, producing a 5°K temperature jump in about 50 μsec. The response of the reacting system is monitored with absorption spectrophotometry.
No takes yet. Share an insight, caveat, or question.
Erman et al. (1966) studied this question.