Key result
Volumetric VH-IVUS shows strong between-centre reproducibility despite minor volume measurement differences.
Observational
Independent analysts
Yes
Answers are grounded in this paper and cite their sources.
AIMS: To assess for the first time in a multicentre design the between-centre reproducibility of volumetric virtual histology intravascular ultrasound (VH-IVUS) measurements with a semi-automated, computerassisted contour detection system in mild-to-moderately diseased coronary segments. METHODS AND RESULTS: Analysts of four European IVUS centres performed independent IVUS analyses (in total 7,188 cross-sectional analyses) and obtained volumetric data to evaluate the reproducibility of volumetric VH-IVUS measurements in 36 coronary segments (length 20.0+/-0.4 mm) from patients with stable angina. Geometric and compositional VH-IVUS measurements were highly correlated for the different comparisons. Overall intraclass correlation for vessel, lumen, plaque volume and plaque burden were 0.98, 0.92 0.95, and 0.86, respectively; for fibrous, fibro-lipidic, necrotic core and calcified volumes overall intraclass correlations were 0.95, 0.93, 0.99, and 1.00, respectively. There were significant but small differences for vessel, lumen, fibrous and calcified volumes, and there was no significant difference for plaque volume. Of the plaque components necrotic core and calcified volume showed on average the highest reproducibility. CONCLUSIONS: These findings underline the necessity to centrally analyse IVUS data obtained in multicentre studies addressing mild-to-moderately diseased coronary arteries. In addition, pooling VH-IVUS data from different studies, analysed at different centres, may be problematical.
No takes yet. Share an insight, caveat, or question.
Huisman et al. (2010) conducted an observational in Mild-to-moderate coronary atherosclerosis (stable angina). Volumetric virtual histology intravascular ultrasound (VH-IVUS) analysis with semi-automated computer-assisted contour detection vs. Between-centre independent analyst comparisons across four European IVUS centres was evaluated on Between-centre reproducibility of volumetric VH-IVUS geometric and compositional measurements (ICC 0.86-1.00 across volumetric parameters). Volumetric VH-IVUS showed high between-centre correlation, though small significant differences existed for vessel, lumen, fibrous, and calcified volumes.
Synapse has enriched 10 related papers. Consider them for comparative context: