Answers are grounded in this paper and cite their sources.
This case report follows a woman who had a total hip replacement in 1992 when she was 45 years old. Six serial computed tomography (CT) examinations over a period of 13 years provided information that allowed her revision surgery to be limited to liner replacement as opposed to replacement of the entire prosthesis. Additionally, they provided data that ruled out the presence of osteolysis and indeed none was found at surgery. In 2004, when the first CT was performed, the 3D distance the femoral head had penetrated into the cup was determined to be 2.6 mm. By 2017, femoral head penetration had progressed to 5.0 mm. The extracted liner showed wear at the thinnest part to be 5.5 mm, as measured with a micrometer. The use of modern CT techniques can identify problems, while still correctable without major surgery. Furthermore, the ability of CT to assess the direction of wear revealed that the liner wear changed from the cranial to dorsal direction.
No takes yet. Share an insight, caveat, or question.
Weidenhielm et al. (2018) studied this question.