Answers are grounded in this paper and cite their sources.
This study reports the fabrication and characterization of a screen-printed carbon electrode (SPCE) modified with pyridinium-antimony trichloride (PAT) for copper ion (Cu²⁺) detection in aqueous environments. The SPCE was prepared by printing carbon ink onto a polyethylene terephthalate-mica substrate, while the PAT material was characterized using transmission electron microscope (TEM), field emission scanning electron microscopy (FESEM), Fourier transform infrared spectroscopy (FTIR), and X-ray diffraction (XRD). Electrochemical performance was evaluated using cyclic voltammetry, electrochemical impedance spectroscopy, and chronocoulometry, exhibited two distinct linear ranges (0.1–10 μM, R2= 0.998; 0.1–10 mM, R2= 0.938). The sensor demonstrated exceptional selectivity against ten interfering metal ions, exhibiting high reproducibility (relative standard deviation, RSD, 2.08%) and repeatability (RSD 3.22%). These metrics confirm the sensor's reliability and consistency in Cu2+detection.
No takes yet. Share an insight, caveat, or question.
Nofriandi et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: