Why the study?
Heart disease is increasing in Indonesia, and cardiovascular care could achieve better diagnosis through continuous monitoring.
The developed Bluetooth-based 12-lead ECG system has limited transmission range, failing at 20 meters without obstacles and at shorter distances when physical barriers are present.
Based on data from Basic Health Research (Riskesdas) in 2018, the incidence of heart and blood vessel disease is increasing from year to year. At least 15 out of 1000 people or about 2,784,064 individuals in Indonesia suffer from heart disease. Therefore, cardiovascular health care can make a better diagnosis through continuous monitoring. The purpose of this study was to develop a 12-lead circuit, a lead selector (Wilson Central Terminal), an instrumentation booster, an analog filter (Notch Filter 50Hz), Arduino UNO, a Bluetooth module, and Delphi7 application to display electrocardiograph signals. The results show that the Bluetooth module cannot send a signal at a distance of 20 meters if there is no obstacle, cannot send a signal at a distance of 10 meters if there is an obstacle in the form of a wall, and cannot send a signal at a distance of 16 meters if there is an obstacle in the form of wood (doors).
No takes yet. Share an insight, caveat, or question.
Mukti et al. (2023) studied this question.