Why the study?
Rapid deployment aortic valves may interfere with the cardiac conduction system, prompting investigation into the need for permanent pacemaker implantation following Edwards INTUITY valve implantation.
What are the rates and predictors of permanent pacemaker implantation following aortic valve replacement with the INTUITY rapid deployment valve in combined procedures?
What are the rates and predictors of permanent pacemaker implantation following aortic valve replacement with the INTUITY rapid deployment valve in combined procedures?
Following aortic valve replacement with the INTUITY rapid deployment valve in combined procedures, 9.4% of patients required a permanent pacemaker, with preoperative RBBB and large valve size identified as independent predictors of this complication.
BACKGROUND: Rapid deployment aortic valves may interfere with the cardiac conduction system. We investigated the need for permanent pacemaker implantation (PPI) following the implantation of Edwards INTUITY valve (Edwards Lifesciences, Irvine, CA). METHODS: One hundred twenty patients underwent aortic valve replacement (AVR) with the INTUITY valve in a combined procedure at the German Heart Centre Munich between April 2016 and December 2019. Twenty-four patients with prior PPI or concomitant ablation procedures (24/120, 20%) were excluded. Patient-specific, procedural and post-procedural outcomes were assessed in the remaining 96 cases. RESULTS: AVR was successful in all cases. Seventy-four percent of the study population were men. Mean age was 69.5±7.6 years. EuroSCORE II was 3.2±2.9. Forty-six patients (46/96, 47.9%) presented with pre-operative conduction disorders, right bundle branch block (RBBB) (17/96, 17.7%) and first-degree or second degree atrio-ventricular block (AVB) (18/96, 18.8%), in particular. In total, 9 patients (9/96, 9.4%) underwent PPI. PPI was required in 3 patients (3/50, 6.0%) who did not have a pre-existing conduction disorder due to new high degree AVB. 6 patients with pre-operative conduction disorders (6/46, 13%) needed PPI. Timing of PPI was 5.2±1.5 days (median 5). Independent predictors of PPI were preoperative RBBB [odds ratio (OR) =4.554, P=0.049] and large valve size (#27) (OR =5.527, P=0.031). CONCLUSIONS: The analysis of the data collected enabled us to identify patient factors associated with higher risk for post-operative PPI following AVR with the INTUITY valve. Patient factors associated with post-operative PPI, were RBBB and large valve size. These patients should be closely monitored following the procedure, in particular.
No takes yet. Share an insight, caveat, or question.
Deutsch et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: