Answers are grounded in this paper and cite their sources.
A high flow rate four-stage impacfor was developed for the determination of aerosol concentration as a function of both particle size and time. The unit is very useful for long-term sampling intervals (24 hr) and for sampling very dusty atmospheres. Gas-borne particulate matter is collected out on four rotating drums, each with a collection surface area of about 10 sq in.; this allows a large quantity of materials to be collected without danger of particle build-up and blow-off. A particle size-collection efficiency calibration for the unit is presented together with experimental data on wall losses, surface coatings, and other important operational considerations.
No takes yet. Share an insight, caveat, or question.
Dale A. Lundgren (1967) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: