The construction and response of the immobilized enzyme chemically modified electrode as an amperometric sensor is described. Glucose oxidase (GOx) has been covalently bound to poly‐2‐aminoaniline (PAANI) film, formed onto a platinized Pt surface via a tetrachloro‐1,4‐ p ‐benzoquinone linkage. The steady‐state amperometric response of the sensor is investigated as a function of the glucose concentration, applied potential and platinum loading. The working potential for glucose determinations was +250mV (vs. SCE) and the biosensor exhibited a linear response in the range 0.2–20 mM with a response time that does not exceed 5 s.