Key result
VAD implantation is linked to acquired von Willebrand syndrome in ~89% of patients vs. none with TAH.
Observational (n=12)
Answers are grounded in this paper and cite their sources.
Unexplained bleeding episodes are associated with ventricular assist devices (VAD) and can occur in part due to acquired von Willebrand syndrome (AVWS). AVWS is characterised by loss of high molecular weight (HMW) multimers of von Willebrand factor (VWF) and decreased ratios of collagen binding capacity and ristocetin cofactor activity to VWF antigen. Loss of multimers can occur as VWF is subjected to increased shear stress, which occurs in presence of VADs. We studied 12 patients who required mechanical support of their native heart for terminal cardiac insufficiency. Nine patients underwent placement of a VAD, while three underwent placement of a total artificial heart (TAH), which is connected directly to heart and large cardiac vessels without cannulas. Within one day of VAD implantation, four of five patients evaluated demonstrated loss of HMW multimers and impaired VWF function. AVWS was present within two weeks of implantation in eight of nine patients, and in all seven tested patients after >/=3 months. Patients with different VAD types developed varying severities of AVWS. After VAD explantation, HMW multimers were detectable and VWF function normalised in all patients. AVWS was not observed in the TAH patients studied. Our findings demonstrate that patients with an implanted VAD experience a rapid onset of AVWS that is quickly and completely reversed after device explantation. In addition, TAH patients do not develop AVWS. These results suggest that shear stress associated with exposure of blood to VAD cannulas and tubes may contribute to the development of AVWS.
No takes yet. Share an insight, caveat, or question.
Beyersdorf et al. (2010) conducted an observational in Terminal cardiac insufficiency (n=12). Ventricular assist device (VAD) implantation vs. Total artificial heart (TAH) implantation was evaluated on Acquired von Willebrand syndrome (AVWS) within two weeks of implantation. Ventricular assist device implantation induced acquired von Willebrand syndrome within 2 weeks in 8 of 9 patients, compared with 0 of 3 total artificial heart recipients.
Synapse has enriched 10 related papers. Consider them for comparative context: