Answers are grounded in this paper and cite their sources.
A quantitative CE (qCE) system with high precision has been developed, in which a 4-port nano-valve was isolated from the electric field and served as sample injector. The accurate amount of sample was introduced into the CE system with high reproducibility. Based on this system, consecutive injections and separations were performed without voltage interruption. Reproducibilities in terms of RSD lower than 0.8% for retention time and 1.7% for peak area were achieved. The effectiveness of the system was demonstrated by the quantitative analysis of caffeine, theobromine, and theophylline in real samples, such as tea leaf, roasted coffee, coca cola, and theophylline tablets.
No takes yet. Share an insight, caveat, or question.
Li et al. (2008) studied this question.