Left bundle branch area pacing after TAVR was complicated by electrical storm with sustained VT at 260 bpm 6 days post-implant, which resolved after lead removal and upgrade to a dual-chamber ICD.
Why the study?
Although TAVR offers faster recovery and lower procedural risk than surgery for aortic stenosis, it carries a higher rate of conduction disturbances and pacemaker implantation.
Case Report (n=1)
No
Answers are grounded in this paper and cite their sources.
Left bundle branch area pacing for complete heart block after TAVR carries a rare risk of life-threatening ventricular proarrhythmia triggered by septal/Purkinje ectopy, which may require lead removal and close post-procedural monitoring.
No abstract yet. The journal hasn't published one we can show.
No takes yet. Share an insight, caveat, or question.
Bhatt et al. (2024) conducted a case report in Intermittent complete heart block following transcatheter aortic valve replacement (TAVR) for severe aortic stenosis (n=1). Left bundle branch area pacing (LBBAP) with a Medtronic SelectSecure 3830 lead was evaluated on Lead-related proarrhythmia (electrical storm / sustained ventricular tachycardia). Left bundle branch area pacing after TAVR was complicated by electrical storm with sustained VT at 260 bpm 6 days post-implant, which resolved after lead removal and upgrade to a dual-chamber ICD.